Colorimetric aluminium Continuous Water Analyser
- Measurement range -0 to 1.5 ppm Al (0 to 1.5 mg/L)
- Chemical Method- Pyrocatechol Violet (PCV)
- Background color correction Compensated at the measurement wavelength
- Measurement frequency-1 to 6 measurements per hour (User-selectable)
- Accuracy / maximum measured error ±5 % of reading or ±0.005 ppm (whichever is the greater)
- Repeatability ±5 % of reading or ±0.005 ppm (whichever is the greater)
- Resolution-0.001 ppm or 1 ppb
- Measurement units mg/l, ppm, ppb, μg/l
- Calibration2-point, automatic calibration

ISE Ammonia Continuous Water Analyser
- Measurement range-0.050 to 1000 ppm NH3 (0.050 to 1000 mg/l)
- Measurement Principle Gas-sensing ammonia probe
- Measurement mode-Continuous
- Accuracy / maximum measured error,±5 % of reading or ±0.02 ppm for 0.05 to 500 ppm (whichever is the greater),±7.5 % of reading for 500 to 1000 ppm
- Repeatability,±2 % of reading or ±0.02 ppm (whichever is the greater)
- Calibration 2-point, automatic calibration
- Measurement units mg/l, ppm, ppb, μg/l

Chlorine Analyser Continuous Water Analyser
- Total and free residual chlorine measurement – add potassium iodide to reagent to convert free chlorine measurement to total chlorine
- Reagentless operation – reduced operating costs with reagentless operation at a pH of 7.5 or less
- Multiple sensor input capability – reduced installation costs for multiple point monitoring operations
- Self-cleaning sensor assembly – prolongs sensor life and maintains accuracy

Conductivity Sensor Continuous Water Analyser
- Stainless steel electrodes enabling measurements down to 0.055 µS/cm ultra-pure water
- Enables installation without the need for costly bypasses
- Enables maintenance without process shutdown
- 1 in. BSP or NPT threaded process connection
- Guaranteed 0.05 and 0.10 cell constants
- Integral Pt100 temperature sensor
- Maximum temperature 110 °C (230 °F)
- Maximum pressure 10.5 bar (150 psi)
- Integral 0.9 m (3 ft) cable

Conductivity Transmitter Continuous Water Analyser
- Transmitter options for all sensor types
— 2-electrode sensors for low conductivity measurement
— 4-electrode sensors for wide range conductivity applications and slurries
— Toroidal conductivity for corrosive and heavy coating applications - Through-the-glass programming (TTG) configuration
- Hazardous area approved- IS/NI and Exd
- Communications- 4 to 20 mA with HART (FDT/DTM/EDD)
- Continuous sensor and self-monitoring- Electronics self-check

Industrial 2-electrode Conductivity Transmitter Continuous Water Analyser
- Annular carbon electrodes enabling measurements up to 10,000 µS/cm
- Screw-in insertion sensor with 1 in. BSP or NPT threads
- 0.10 and 1.0 cell constants
- Integral Pt100 temperature sensor
- Maximum temperature 100 °C (212 °F)
- Maximum pressure 7 bar (100 psi)
- Ingress protection IP 68
- Fixed cable, terminal head or detachable cable versions

pH/Redox Continuous Water Analyser
- User-customizable for insertion, f-through, immersion and hot-tap retractor
- Economical investment, powerful performance
- Combination Sensor complete with measuring, reference and temperature elements
- Bulb and flat glass options
- High stability PTFE reference junction
- Gel-filled disposable sensor
- Process-resistant gel electrolyte

Multiwave Photometer Continuous Gas Analyser
- The Multiwave Photometer PFO3372 incorporates the use of fiber optic sampling accessories for added measurement capability.
- The ABB Multiwave photometers provide continuous measurements in complex streams which allow for fast reactions to process changes.
- The fast analysis provides optimum control of a process and improves its safe operation.
- Measurements can be made in liquid or vapor streams. Simple installation on wall or tank.
<?h6>

Zirconia In-Situ Oxygen & Combustion Continuous Gas Analyser
- Oxygen only or oxygen plus combustibles
– increased combustion efficiency
– burner malfunction identification
– enhanced plant safety - Comprehensive diagnostics
– NAMUR-compliant diagnostic symbols
– supports predictive maintenance
– fully logged diagnostic events - Process logging and trending of all measured and
calculated values
– oxygen and carbon monoxide equivalent (COe)
– process temperature measurement
– combustion efficiency calculation - Multiple sample filter and blowback options, optional dual filtration system, fully programmable blowback.

Laser Process Analyser
- Real-time continuous on-line measurements of several gases simultaneously
- Gases measured include CH4, C2H2, CO, CO2, H2O, H2S, HCl, HF, NH3, O2, and others
- User selectable data rate
- Industry Standard Communications, including analog 4 to 20 mA and digital Modbus
- Hazloc-approved USB portal access for data file download and simple field service

Turbidity Meter Continuous Water Analyser
- Range -Turbidity: 0 to 4000 NTU,Suspended solids: 0 to 100,000 mg/l SiO2 (100 g/l)
- Accuracy-Turbidity: <±2 % measured value,Suspended solids: dependent on sample
- Method-90° scattered light measurement in accordance with DIN/EN 27027/ISO 7027
- Repeatability,<1 %
- Limit of detection-0.006 NTU
- Response time-T90 < 30 s with filtering disabled
- Storage conditions–5 to 70 °C (23 to 158 °F)
- Operating temperature-0 to 60 °C (32 to 140 °F)
- Operating pressure-Up to 10 bar (145 psi) for metal versions